Table T7. Analytical conditions for ICP-AES analyses.
| Element | Wavelength (nm) | Integration time per calculation point (s) | Photomultiplier voltage (V) | Photomultiplier gain | Increment between points (nm) | Number of calculation points | Limit of detection (ppm) | 
|---|---|---|---|---|---|---|---|
| Al† | 396.152 | 0.5 | 720 | 1 | 0.002 | 7 | 0.109 | 
| Ba | 455.403 | 0.5 | 730 | 10 | 0.002 | 7 | 10 | 
| Ca† | 393.366 | 0.5 | 350 | 1 | 0.002 | 7 | 0.084 | 
| Co | 228.616 | 0.5 | 990 | 100 | 0.002 | 7 | 39 | 
| Cr | 267.716 | 0.5 | 860 | 100 | 0.002 | 7 | 22 | 
| Cu | 324.754 | 0.5 | 690 | 10 | 0.002 | 7 | 112 | 
| Fe† | 259.940 | 0.5 | 560 | 10 | 0.002 | 7 | 0.328 | 
| K† | 766.490 | 0.5 | 990 | 100 | 0.002 | 7 | 0.014 | 
| Mg† | 285.213 | 0.5 | 600 | 10 | 0.002 | 7 | 0.220 | 
| Mn† | 257.610 | 0.5 | 580 | 100 | 0.002 | 7 | 0.016 | 
| Na† | 589.592 | 0.5 | 610 | 10 | 0.002 | 7 | 0.086 | 
| Ni | 231.604 | 0.5 | 990 | 100 | 0.002 | 7 | 55 | 
| P† | 178.229 | 0.5 | 990 | 100 | 0.002 | 7 | 0.143 | 
| Si† | 251.611 | 0.5 | 570 | 10 | 0.002 | 7 | 0.522 | 
| Sr | 407.771 | 0.5 | 560 | 10 | 0.002 | 7 | 2 | 
| V | 292.402 | 0.5 | 960 | 100 | 0.002 | 7 | 3 | 
| Y | 371.029 | 0.5 | 620 | 100 | 0.002 | 7 | 2 | 
| Zr | 343.823 | 0.5 | 640 | 100 | 0.002 | 7 | 6 | 
| Ti† | 334.941 | 0.5 | 570 | 10 | 0.002 | 7 | 0.021 | 
| Sc | 361.384 | 0.5 | 660 | 100 | 0.002 | 7 | 1 | 
Notes: ICP-AES = inductively coupled plasma–atomic emission spectrometry. † = Limit of detection reported as oxide weight percent. Detection limits of the instrument represent three times the instrumental blank (See “Analysis” in “ICP-AES analyses of major and trace elements”).