Table AT1. Temperature calibration of DVTPP. A. Calibration coefficients for data loggers and thermistors. Channel T1 is the logger channel used to measure the thermistor. Rt′ = thermistor resistance in ohms. B. Actual thermistor calibration data.
Table AT1A. Calibration coefficients. (See table notes.)
| Calibration coefficient |
Logger 9367 | Logger 9368 |
|---|---|---|
| R1 | 251680.977 | 249507.092 |
| k0 | 65564.0623 | 65534.6958 |
| k1 | 65550.3036 | 65533.6013 |
| Calibration coefficient |
Thermistor 0226-3 | Thermistor 0226-2 |
|---|---|---|
| A | 4.60148156E–04 | 4.52826700E–04 |
| B | 2.10947147E–04 | 2.11217111E–04 |
| C | 6.41209309E–08 | 6.19876025E–08 |
Notes: For Channel T1 calibration, Rt′ = R1 × (k1 – x)/[k0 – (k1 – x)], where x = T1 counts. For thermistor calibration (ohms to Kelvin [Steinhart and Hart]), 1/t = A + B × Ln(Rt′) + C × Ln(Rt′)3, where °C = K – 273.15.
Table AT1B. Temperature vs. resistance.
| Serial number |
Resistance (Ω) | |||
|---|---|---|---|---|
| 0.000°C | 50.000°C | 100.000°C | 150.000°C | |
| 0226-2 | 1666200 | 162749 | 27716 | 6963 |
| 0226-3 | 1602700 | 157535 | 26907 | 6781 |
Close | Next | Top of page